CASE REPORT
Allergic Response to Pace Maker with Non Healing Wound
 
More details
Hide details
1
SKIMS, Soura, Srinagar, Kashmir
 
2
Cardiology SMHS and associated hospitals, Srinagar, Kashmir
 
 
Publication date: 2014-10-15
 
 
Corresponding author
Imran Hafeez   

Dept. of Cardiology, SKIMS Srinagar, Kashmir, J&K 190011
 
 
Eur J Gen Med 2014;11(4):290-292
 
KEYWORDS
ABSTRACT
Allergic reactions to pacemaker components are rare but well known . We present a case who developed pain, erythema and serosangious discharge immediately after permanent pace maker (PPM) implantation. There was no evidence of infection from ancillary investigations and the whole pacemaker system was explanted. New system was implanted on the opposite side and the patient once again had erythema and swelling of the wound with scant discharge. Hemogram revealed Eosinophillia and also IgE levels were increased. Patient was started on steroids which resulted in prompt healing of wound.